Intermediate Disinfection
To disinfect your scanner (Intermediate):
- Make sure the fan is still detached from the scanner.
- Disinfect the scanner by wiping it with a cloth moistened with a compatible disinfectant.
Alternatively, use a premoistened disinfectant wipe. Use a swab for unreachable areas. - Air-dry. Alternatively, use a non-linting cloth.
- Examine the scanner for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the scanner and contact Clarius Mobile Health.
High-Level Disinfection
- Make sure the fan (and battery on 1st gen and HD models) is detached from the scanner. It is important to disinfect each piece individually.
- Mix the disinfectant solution by following the disinfectant label instructions for solution strength and disinfectant contact duration.
- Using a compatible disinfectant at a temperature of 23°C (73°F), immerse the scanner (and the battery* on 1st gen and HD models) in the disinfectant solution for the manufacturer recommended period of time. If your disinfectant manufacturer does not have a specific period of time listed on the product label, please contact the disinfectant manufacturer for more details.
- Using the instructions on the disinfectant label, rinse the scanner (and battery on 1st gen and HD models).
- Air-dry each piece. Alternatively, use a non-linting cloth.
- Examine the components for damage, such as cracks or splitting where fluid can enter. If damage is evident, do not use the scanner and contact Clarius Mobile Health
Disinfecting the Clarius Fan (HD3 Power Fan and HD)
Before you begin disinfecting, make sure you have cleaned the fan.
Because the fan cannot be submerged in liquid, never use high-level disinfectants. Always use intermediate-level disinfection. For a list of disinfectants recommended for intermediate
disinfection of the Fan, refer to Cleaners & Disinfectants article or from the User Manual.
- Make sure that the fan is detached from the scanner.
- Disinfect the fan by wiping it with a cloth moistened with a compatible disinfectant. Alternatively, use a pre-moistened disinfectant wipe.
- Air-dry. Alternatively, use a non-linting cloth.
- Examine the fan for damage, such as cracks or splitting. If damage is evident, do not use the fan and contact Clarius Mobile Health.
Disinfecting the Micro USB Cable
Before you begin disinfecting, make sure you have cleaned the micro USB cable.
Because the cable cannot be submerged in liquid, never use high-level disinfectants. Always use intermediate-level disinfection. For a list of disinfectants recommended for intermediate
disinfection of the Charger, refer to Cleaners & Disinfectants article or from the User Manual.
- Make sure the Micro USB cable is detached from the power supply and the Clarius
Power Fan HD3. - Disinfect the Micro USB cable by wiping it with a cloth moistened with a compatible
disinfectant. Alternatively, use a premoistened disinfectant wipe. - Air-dry. Alternatively, use a non-linting cloth.
- Examine for damages, such as cracks or splits. If damage is evident, do not use the
Micro USB cable and contact Clarius Mobile Health
Disinfecting the Charger
Before you begin disinfecting, make sure you have cleaned the Charger.
Because the charger cannot be submerged in liquid, never use high-level disinfectants. Always use intermediate-level disinfection. For a list of disinfectants recommended for intermediate
disinfection of the Charger, refer to Cleaners & Disinfectants article or from the User Manual.
- Disinfect the Charger by wiping it with a cloth moistened with a compatible disinfectant. Alternatively, use a premoistened disinfectant wipe.
- Air-dry. Alternatively, use a non-linting cloth.
- Examine for damages, such as cracks or splits. If damage is evident, do not use the Charger and contact Clarius Mobile Health.
*If the battery has been recently used or charged, wait 30 seconds before submerging it in any liquid.